A Menu Structure FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 181
input of the pressure for the local standard con-
ditions
range: 0.70001...1.30000 bar
input of the temperature for the local standard
conditions
range: -20.0...+90.0 °C
SYSTEM settings\Proc. outputs
selection of the displays for the setting of the
outputs of the transmitter
selection of the output to be installed
SYSTEM settings\Storing
selection of the displays for the storing of mea-
sured values in the data logger
setting of the overflow behavior of the data log-
ger
x
selection of the sample mode
• sample: storing and online transmission of
the displayed measured value
• average: storing and online transmission of
the average of all measured values of a stor-
age interval
x
setting of the storing behavior of the totalizers
• one: the value of the totalizer that is currently
displayed will be stored
• both: one value for each flow direction will be
stored
x
cold start
resistant
Normal pressure
1.01325 bar
SYSTEM settings ↕
Proc. outputs
Install Output ↕
Current I1
SYSTEM settings ↕
Storing
Storage mode
sample >AVERAGE<
Quantity Storage
one >BOTH<