180 UMFLUXUS_G6V4-2EN, 2011-03-29
FLUXUS G60x A Menu Structure
Input of the cut-off flow for negative measured
values
range: -12.7...0 cm/s
default: -2.5 cm/s
This display will only be indicated if Cut-off
Flow = sign und Cut-off Flow = user has
been selected.
Input of the cut-off flow for the absolute value of
the measured values
range: 0...12.7 cm/s
default: 2.5 cm/s
This display will only be indicated if Cut-off
Flow = absolut und Cut-off Flow = user
has been selected.
input of an upper limit of the flow velocity
range: 0.1...25.5 m/s
All measured values that are greater than the
limit will be marked as outliers.
Input of 0 (zero) switches off the detection for
outliers.
x
activation of the overflow of the totalizers x
activation of the taking-over of the totalizer val-
ues after a restart of the measurement
x
SYSTEM settings\Gas-Measuring
selection of the displays for the input of the
standard conditions for the gas measurement
This display will only be indicated if Special
Funct.\SYSTEM settings\Measuring\
Gas-Measuring is activated.
cold start
resistant
SYSTEM settings ↕
Gas-Measuring