A Menu Structure FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 179
assignment of temperature inputs and other in-
puts to the measuring channels
SYSTEM settings\Measuring
selection of the displays for the settings of the
measurement
activation of the gas measurement
default: on
activation of the display for the difference be-
tween the measured and the expected sound
speed of a selected reference medium during
the measurement
selection whether the flow velocity is displayed
and transmitted with or without profile correction
x
selection of the input of a lower limit for the flow
velocity:
• absolut: independent of the flow direction
• sign: dependent on the flow direction
x
activation of the input of a lower limit of the flow
velocity:
• factory: the default limit of 2.5 cm/s will be
used
• user: input of a limit
input of the cut-off flow for positive measured
values
range: 0...12.7 cm/s (0.127 m/s), default: 2.5
cm/s (0.025 m/s)
This display will only be indicated if Cut-off
Flow = sign and Cut-off Flow = user has
been selected.
cold start
resistant
Proc. inputs ↕
Link temperature
SYSTEM settings ↕
Measuring
Flow Velocity
normal >UNCORR.<
Cut-off Flow
absolut >SIGN<
Cut-off Flow
factory >USER<