B Technical Data FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 211
Lamb Wave Transducers (zone 1)
technical type GRG1NW3 GTG1NW3 GRH1NW3 GTH1NW3 GRK1NW3 GTK1NW3
order code GLG-
NA1NL
GLG-
NA1NL/LC
GLH-
NA1NL
GLH-
NA1NL/LC
GLK-
NA1NL
GLK-
NA1NL/LC
transducer fre-
quency
MHz 0.2 0.3 0.5
medium pressure
1
min. extended bar metal pipe: 10 metal pipe: 10 metal pipe:
10 (d > 120 mm),
5 (d < 120 mm)
min. bar metal pipe: 15
plastic pipe: 1
metal pipe: 15
plastic pipe: 1
metal pipe:
15 (d > 120 mm),
10 (d < 120 mm)
plastic pipe: 1
inner pipe diameter d
min. extended mm 190 120 60
min. recom-
mended
mm 220 140 80
max. recom-
mended
mm 900 600 300
max. extended mm 1600 1000 500
pipe wall thickness
min. mm 11 7 4
max. mm 23 15 9
material
housing PPSU
with stainless
steel cap and
transducer shoe
304
(1.4301)
PPSU
with stainless
steel cap and
transducer shoe
304
(1.4301)
PPSU
with stainless
steel cap and
transducer shoe
304
(1.4301)
contact surface PPSU PPSU PPSU
degree of protec-
tion according to
IEC/EN 60529
IP65 IP65 IP65
transducer cable
type 1699 1699 1699 1699 1699 1699
length m 5 9 5 9 5 9
dimensions
length l mm 136.5 136.5 136.5
width b mm 59 59 59
height h mm 90.5 90.5 90.5
dimensional draw-
ing
operating temperature
min. °C -40 -40 -40
max. °C +170 +170 +170
temperature
compensation
xxx
l
h
b
l
h
b
l
h
b

Table of Contents