6 Mounting
6.2 Transducers FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
42
Step 1: Disassembly of the transducer mounting fixture Variofix L
• Disassemble the transducer mounting fixture Variofix L.
Step 2: Fixation of the clasps to the tension straps
• Select the installation instruction of the supplied clasp:
Band clamp clasp
The clasp is fixed to the tension strap (see Fig. 6.15).
Quick release clasp
The clasp is fixed to the tension strap (see Fig. 6.16).
• Cut the tension strap to length (pipe circumference + at least 120 mm).
Fig. 6.14: Disassembly of the transducer mounting fixture Variofix L
1 – cover
2 – screw
3 – nut
4 – rail
5 – tension strap clamp
Fig. 6.15: Band clamp clasp with tension strap Fig. 6.16: Quick release clasp with tension strap
1
2
3
4
5
5