7 Connection
FLUXUS G801 7.1 Transducers
63
UMFLUXUS_G801V1-0EN, 2020-06-26
7.1.3 Connection of the transducer cable to the junction box
7.1.3.1 Transducer cable with plastic jacket and stripped ends
Remove the blind plug for the connection of the transducer cable.
Open the cable gland of the transducer cable. The compression part remains in the cap nut.
Push the transducer cable through the cap nut and the compression part.
Prepare the transducer cable.
Shorten the external shield and brush it back over the compression part.
Screw the gasket ring side of the basic part into the junction box.
Insert the transducer cable into the junction box.
Fix the cable gland by screwing the cap nut onto the basic part.
Connect the transducer cable to the terminals of the junction box.
FLUXUS *801C24, *801**-A10****-FF
Notice!
For good electromagnetic compatibility (EMC), it is important to ensure good electrical contact between the external
shield and the cap nut (and thus the housing).
Tab. 7.4: Terminal assignment
terminal connection
V transducer (core)
VS transducer (internal shield)
RS transducer (internal shield)
R transducer (core)
Fig. 7.6: Connection of the extension cable to the transmitter
1 cap nut
2 compression part
3 basic part
4 external shield, brushed back
5 extension cable
cable gland
1 2 3
70 mm
20 mm
10 mm
4 5