7 Connection
7.1 Transducers FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
62
7.1.2 Connection of the extension cable to the transmitter
• The extension cable is connected to the transmitter via the transducer connection.
• Remove the blind plug for the connection of the transducer cable.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut and the compression part.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the basic part into the transmitter housing.
• Insert the extension cable into the housing.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the transmitter.
Notice!
For good electromagnetic compatibility (EMC), it is important to ensure good electrical contact between the external
shield and the cap nut (and thus the housing).
Tab. 7.3: Terminal assignment
terminal connection
AV white or marked cable (core)
AVS white or marked cable (internal shield)
ARS brown cable (internal shield)
AR brown cable (core)
Fig. 7.6: Connection of the extension cable to the transmitter
FLUXUS *801**-A10****-AA, *801**-A10****-AP
1 – cap nut
2 – compression part
3 – basic part
4 – external shield, brushed back
5 – extension cable
70 mm
20 mm
10 mm
4 5
cable gland
1 2 3