9 Measurement
FLUXUS G801 9.1 Parameter input
87
UMFLUXUS_G801V1-0EN, 2020-06-26
Fluid pressure
The fluid pressure is used for the interpolation of the sound speed and the gas compressibility coefficient.
• Enter the fluid pressure.
• Press ENTER.
This display will only be indicated if the menu item Special Funct.\SYSTEM settings\Measuring\Gas-Measuring
is activated OR if it is deactivated and the menu item Special Funct.\SYSTEM settings\Dialogs/Menus\
Fluid pressure is activated.
9.1.3 Other parameters
Transducer parameter
If transducers are identified at a measuring channel, the input of parameters is finished. Press ENTER. The main menu
will be displayed.
If no or special transducers are connected, the transducer parameters have to be entered.
• Select Standard to use the standard transducer parameters stored in the transmitter.
• Select Special Version to enter the transducer parameters. The transducer parameters have to be provided by FLEXIM.
• Press ENTER.
If Special Version is selected, enter the 6 transducer parameters specified by FLEXIM. Press ENTER after each input.
Extension cable
In case the transducer cable is extended (e.g., between junction box and transmitter), enter the length of the extension cable.
• Select the menu item Additional cable in the program branch Parameter.
• Enter the length of the extension cable.
• Press ENTER.
Parameter\Fluid pressure
Parameter\Transducer Type
Notice!
If a standard transducer is selected, no transducer-specific calibration values are considered. A higher uncertainty
has to be expected.
Parameter\Transducer Type\Special Version
Parameter\Additional cable