9 Measurement
9.2 Measurement settings FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
92
9.2.3.1 Output of the measured value
Output range
When configuring an analog output, the output range has
to be defined.
• Select the menu item I1 Output range.
• Press ENTER.
• Select a list item.
– 4/20 mA
– other range
• Press ENTER.
• If other range is selected, enter the values Output MIN and Output MAX.
• Press ENTER after each input.
Error output
In the following dialog, an error value can be defined which is to be transmitted if the source item cannot be measured,
e.g., if there are solids in the fluid.
• Select a list item for the error output.
• Press ENTER.
• If Other value is selected, enter an error value. The value has to be within the output range.
• Press ENTER.
Source item
Output range
Error-value
active loop
Output Test
see annex A
Special Funct.\SYSTEM settings\
Proc. outputs\...\I1 Output range
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Error-value
Notice!
The settings will be stored at the end of the dialog.
Tab. 9.2: Error output
error value description
Minimum The lower limit of the output range is output.
Hold last value The last measured value is output.
Maximum The upper limit of the output range is output.
Other value The value has to be entered manually. It has to be within the limits of the output.
Example
source item:
output:
output range:
error delay:
volumetric flow rate
current
4…20 mA
t
d
> 0
The volumetric flow rate cannot be measured during the
time interval t
0
...t
1
. The error value will be output.
v [m³/h]
t
0
t
1
t