9 Measurement
9.2 Measurement settings FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
94
Output function test
The function of the output can now be tested.
• Connect a measuring instrument to the terminals of the installed output.
• Enter a test value. It has to be within the output range.
• Press ENTER.
If the measuring instrument displays the entered value, the output functions correctly.
• Select yes to repeat the test, or no to return to the menu item SYSTEM settings.
• Press ENTER.
9.2.4 Activation of an analog output
• Select the channel for which an output is to be activated
in the program branch Output Options.
• Press ENTER.
This display will not be indicated if the transmitter has only
one measuring channel.
• Press ENTER until Current Loop is displayed.
• Select yes to activate the output.
• Press ENTER.
Measuring range
After an analog output has been activated in the program branch Output Options, the measuring range of the source
item has to be entered.
• Select sign if the sign of the measured values is to be considered for the output.
• Select absolut if the sign of the measured values is not to be considered for the output.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Output Test
Special Funct.\SYSTEM settings\Proc. outputs\...\I1= 10 mA\Again?
Output Options
for Channel x
ENTER until
Current Loop
Meas.Values
Zero-Scale Val.
Full-Scale Val.
Test output
Error-val. delay
see annex A
Notice!
An output can only be activated in the program branch
Output Options if it has previously been installed.
Output Options\for Channel A
Output Options\...\Current Loop
Output Options\...\Meas.Values