9 Measurement
FLUXUS G801 9.2 Measurement settings
95
UMFLUXUS_G801V1-0EN, 2020-06-26
• Enter the lowest expected measured value. The unit of measurement of the source item will be displayed.
Zero-Scale Val. is the value assigned to the value Output MIN of the output range.
• Press ENTER.
• Enter the highest expected measured value. The unit of measurement of the source item will be displayed.
Full-Scale Val. is the value assigned to the value Output MAX of the output range.
• Press ENTER.
Function test
The function of the output can now be tested.
• Connect a measuring instrument to the terminals of the installed output.
• Select yes to test the output.
• Press ENTER.
• Enter a test value for the selected physical quantity. If the measuring instrument displays the entered value, the output
functions correctly.
• Press ENTER.
• Select yes to repeat the test.
• Press ENTER.
Output Options\...\Zero-Scale Val.
Output Options\...\Full-Scale Val.
Example
output: current
output range: 4…20 mA
Zero-Scale Val.: 0 m³/h
Full-Scale Val.: 300 m³/h
volumetric flow rate = 0 m³/h, corresponds to 4 mA
volumetric flow rate = 300 m³/h, corresponds to 20 mA
Output Options\...\I1:Test output?
Output Options\...\I1:Test value
Output Options\...\I1:Test output?
Example
output: current
output range: 4…20 mA
Zero-Scale Val.: 0 m³/h
Full-Scale Val.: 300 m³/h
Test value: 150 m³/h (center of the measuring range, corresponds to 12 mA)
If the measuring instrument displays 12 mA, the current output functions correctly.