9 Measurement
FLUXUS G801 9.2 Measurement settings
95
UMFLUXUS_G801V1-0EN, 2020-06-26
Enter the lowest expected measured value. The unit of measurement of the source item will be displayed.
Zero-Scale Val. is the value assigned to the value Output MIN of the output range.
Press ENTER.
Enter the highest expected measured value. The unit of measurement of the source item will be displayed.
Full-Scale Val. is the value assigned to the value Output MAX of the output range.
Press ENTER.
Function test
The function of the output can now be tested.
Connect a measuring instrument to the terminals of the installed output.
Select yes to test the output.
Press ENTER.
Enter a test value for the selected physical quantity. If the measuring instrument displays the entered value, the output
functions correctly.
Press ENTER.
Select yes to repeat the test.
Press ENTER.
Output Options\...\Zero-Scale Val.
Output Options\...\Full-Scale Val.
Example
output: current
output range: 4…20 mA
Zero-Scale Val.: 0 m³/h
Full-Scale Val.: 300 /h
volumetric flow rate = 0 m³/h, corresponds to 4 mA
volumetric flow rate = 300 m³/h, corresponds to 20 mA
Output Options\...\I1:Test output?
Output Options\...\I1:Test value
Output Options\...\I1:Test output?
Example
output: current
output range: 4…20 mA
Zero-Scale Val.: 0 m³/h
Full-Scale Val.: 300 /h
Test value: 150 m³/h (center of the measuring range, corresponds to 12 mA)
If the measuring instrument displays 12 mA, the current output functions correctly.