6 Installation
FLUXUS G809 6.2 Transducers
51
UMFLUXUS_G809V2-2EN, 2020-06-25
Insert approx. 100 mm of the tension strap into one of the slots of the tension strap clamp and bend it, see Fig. 6.40.
If necessary, insert the long end of the tension strap into the metal spring, see Fig. 6.41. The metal spring does not have
to be mounted on:
steel pipes
pipes with an outer pipe diameter < 80 mm
pipes that are not subjected to significant temperature fluctuations
Place the tension strap around the pipe, see Fig. 6.42.
Position the tension strap clamp and metal spring (if necessary) on the pipe, see Fig. 6.42:
on a horizontal pipe, mount the tension strap clamp on the side of the pipe, if possible
mount the metal spring (if necessary) opposite to the tension strap clamp
Insert the long end of the tension strap into the second slot of the tension strap clamp, see Fig. 6.42 a.
Tighten the tension strap and bend it.
Bend both ends of the tension strap, see Fig. 6.42 b.
Repeat the steps for the second tension strap. Position the tension strap at the distance s, see Fig. 6.42.
Put the rail on the tension strap clamps.
Use the screws to fix the rail to the tension strap clamps, see Fig. 6.43.
Tighten the screws.
Fig. 6.40: Tension strap with clamp Fig. 6.41: Tension strap with metal spring and clamp
1 tension strap clamp 1 tension strap clamp
2 metal spring
Fig. 6.42: Tension strap with tension strap clap and metal spring on the pipe
1 metal spring
2 tension strap clamp
s = length of the rail - 33 mm
Fig. 6.43: Rail on the pipe
1 screws
1 1
2
1 2 a
b
s
1