EasyManua.ls Logo

Fluke 1560 Black Stack - Calibration Procedure (2560)

Fluke 1560 Black Stack
187 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SELECT A DEVICE TO CALIBRATE
>SPRT 1
TCS 2
PRTS 3
After the device is selected a new window appears showing the parameters and
functions available from the device. New values can be entered for the parame
-
ters using the numeric buttons and pressing ENTER.TheUD buttons can be
used to move between parameters.
CALIBRATE DEVICE: SPRT
0 ADJ: 0.0
100 ADJ: 0.0
400 ADJ: 0.0
CAL DATE: 05-21-96
SER NUM: A26123
For the 2560/2567 SPRT module the list of parameters includes 0_ADJ,
100_ADJ,and400_ADJ as described above. The list also includes the calibra
-
tion date (CAL DATE) parameter, used to record the date the module was cali
-
brated, and the (SER NUM) parameter, used to record the serial number of the
module.
7.4.3 Calibration Procedure (2560)
Calibration requires adjustment of the 0_ADJ, 100_ADJ,and400_ADJ parame
-
ters at three specific input resistances. If the resistances used are approximately
0Ω, 100Ω and 400Ω respectively the adjustments are independent and the pro
-
cedure is simple. The order in which the adjustments are performed is impor
-
tant. The adjustment of the 400_ADJ parameter must be performed last as the
adjustments of 0_ADJ and 100_ADJ affect the measurement at 400Ω but
400_ADJ does not affect the measurements at 0Ω and 100Ω. Either channel
can be used for calibration. Set the conversion type to R(Ω)todisplayresis
-
tance (see Section 5.2.1.1 ). The accuracy required of the resistance standards is
1/4 of the instrument accuracy: that is ±0.00012Ω at 0Ω, ±0.0005Ω (5 ppm) at
100Ω, and ±0.002Ω (5 ppm) at 400Ω. The recommended procedure is as
follows:
117
7 2560/2567 SPRT Module
Calibration

Table of Contents

Related product manuals