6.5.4.1 CALCulate[n]:CONVert:CATalog?
This query command returns a list of conversion types available for the given
input channel. The channel number is given by the CALC suffix n.There
-
sponse list gives the names of the conversion types as strings (enclosed in dou
-
ble quotes) separated by commas. The available conversion types depend on the
type of module to which the channel belongs. Table 14 lists the names of con
-
version types available with each input module type.
Example command: CAL2:CONV:CAT?
Example response:
“I90”, ”RES”,”W”,”I68”,”CVD”,”POLY”
Input Type Conversion types Mnemonic
PRT/SPRT ITS-90 (default) I90
R(
Ω
)
RES
W(T90) W
IPTS-68 I68
CVD CVD
POLYNOMIAL POLY
Thermistor THRM-R(T) (default) TRES
R(
Ω
)
RES
THRM-T(R) TTEM
POLYNOMIAL POLY
2564 module ITS-90 I90
R(
Ω
)
RES
W(T90) W
ITPS-68 I68
CVD CVD
POLYNOMIAL POLY
THRM-R(T) TRES
THRM-T(R) TTEM
Thermocouple TC-K (default) K
VOLTS VOLT
TC-B B
TC-E E
TC-J J
1560 Thermometer Readout
User’s Guide
82
Table 14 Conversion Type Mnemonics

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1560 Black Stack and is the answer not in the manual?

Fluke 1560 Black Stack Specifications

General IconGeneral
BrandFluke
Model1560 Black Stack
CategoryThermometer
LanguageEnglish

Related product manuals