6.5.4.1 CALCulate[n]:CONVert:CATalog?
This query command returns a list of conversion types available for the given
input channel. The channel number is given by the CALC suffix n.There
-
sponse list gives the names of the conversion types as strings (enclosed in dou
-
ble quotes) separated by commas. The available conversion types depend on the
type of module to which the channel belongs. Table 14 lists the names of con
-
version types available with each input module type.
Example command: CAL2:CONV:CAT?
Example response:
“I90”, ”RES”,”W”,”I68”,”CVD”,”POLY”
Input Type Conversion types Mnemonic
PRT/SPRT ITS-90 (default) I90
R(
Ω
)
RES
W(T90) W
IPTS-68 I68
CVD CVD
POLYNOMIAL POLY
Thermistor THRM-R(T) (default) TRES
R(
Ω
)
RES
THRM-T(R) TTEM
POLYNOMIAL POLY
2564 module ITS-90 I90
R(
Ω
)
RES
W(T90) W
ITPS-68 I68
CVD CVD
POLYNOMIAL POLY
THRM-R(T) TRES
THRM-T(R) TTEM
Thermocouple TC-K (default) K
VOLTS VOLT
TC-B B
TC-E E
TC-J J