EasyManua.ls Logo

Fluke 1560 Black Stack - Table 30 3560 Specifications

Fluke 1560 Black Stack
187 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1560 Thermometer Readout
User’s Guide
164
GPIB (IEEE-488) interface
Capability IEEE-488.2, 1992
SH1, complete source handshake capability
AH1, complete acceptor handshake capability
T6, basic talker, serial poll, no talk-only capability, unaddress if MLA
TE0, no extended-talk capability
L4, basic listener, no listen-only capability, unaddress if MTA
LE0, no extended-listen capability
SR1, complete serial poll capability
RL0, no remote-local capability
PP0, no parallel poll capability
DC1, complete device clear capability
DT0, no device trigger capability
C0, no control capability
E2, three-state drivers
Parallel (Centronics) printer interface
Connector 25-pin subminiature D
Serial (RS-232) interface
Connector 9-pin subminiature D
baud rate 1200 to 19200
Protocol 8 data bits, 1 stop bit, no parity
Analog output
Number of channels 1
Connector two-terminal screw clamp
Voltage range –1.25 to +1.25V
Voltage resolution 0.0006V
Digital resolution 12 bits
Maximum source resistance
1
Ω
Maximum output current 2 mA
Linearity ±0.0006V
Typical accuracy
±2% ±0.02V
Isolation grounded
Recommended operating temperature range 18 to 28°C (64 to 82°F)
Absolute operating temperature range 5 to 35°C (40 to 95°F)
Weight 2.2 lbs. (1 kg.)
Accuracy of the analog output can be significantly improved by calibration
Table 30 3560 Specifications