EasyManua.ls Logo

Fluke 1560 Black Stack - 2562;2568 PRT Scanner Module; Description; Specifications

Fluke 1560 Black Stack
187 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 2562/2568 PRT Scanner Module
This section explains the features and operation of the optional 2562/2568 PRT
Scanner Module.
9.1 Description
The 2562/2568 PRT Scanner Module is an add-on module that allows the 1560
to measure temperature with up to eight SPRT, PRT, and RTD sensors. The
modules accept two-, three-, or four-wire sensors with nominal resistance of 25
and 100Ω (2562) or 500 and 1KΩ (2568). The excitation current alternates po
-
larity to minimize the effects of thermoelectric offsets. Up to eight sensors can
be connected to the 2562/2568 simultaneously and measured alternately.
9.2 Specifications
2562 2568
Resistance range
0 to 400
Ω
0 to 4 k
Ω
Resistance accuracy, one-year
(Using four-wire connection)
1, 3
0 to 25
Ω
: 0.001
Ω
25 to 400
Ω
: 40 ppm of read-
ing
0 to 250
Ω
: 0.01
Ω
250 to 4K
Ω
: 40 ppm of read-
ing
Resistance accuracy, short-term
(Using four-wire connection)
1, 2, 3
0 to 25
Ω
: 0.0005
Ω
25 to 400
Ω
: 20 ppm of read-
ing
0 to 250
Ω
: 0.0075
Ω
250 to 4K
Ω
: 30 ppm of read-
ing
Temperature accuracy, one-year (Using
four-wire connection, not including probe
uncertainty)
1, 3
–100°C
0°C
100°C
200°C
300°C
400°C
500°C
600°C
±0.006°C
±0.010°C
±0.014°C
±0.020°C
±0.024°C
±0.028°C
±0.034°C
±0.040°C
Temperature coefficient
1
0.5 ppm/°C 2.5 ppm/°C
Excitation current 1.0, 1.4 mA; 1 Hz 0.1, 0.05 mA; 1 Hz
Maximum lead resistance
100
Ω
Sample time 2 seconds
Number of channels 8
Recommended operating temperature
range
1
18 to 28°C (64 to 82°F)
Absolute operating temperature range 5 to 35°C (40 to 95°F)
Weight 2.5 lbs. (1.1 kg)
125