5.2.3 Test Conversion
The TEST CONV soft-key allows you to test the probe characterization algo
-
rithm and characterization coefficients for a specific probe. You must first se
-
lect the channel number of the probe. A window appears requesting you to
select the input channel. Each channel in the list is identified with the device
name, the channel number, and the probe serial number. The channel is selected
using the UD buttons and pressing ENTER.
SELECT A PROBE TO TEST
>SPRT 01, PROBE 566-011
SPRT 02, PROBE 566-012
TCS 03, PROBE 1341
TCS 04, PROBE 1342
TCS 05, PROBE 1343
TCS 06, PROBE 1344
$
After the probe is selected, a new window appears allowing you to test the tem-
perature calculation for the probe. You can enter an input value (e.g. resistance)
and the corresponding output value (e.g. temperature) will be shown. Press
EXIT to exit.
TEST CONVERSION
RESISTANCE: 100.0145
TEMP (C): 0.0101
5.3 Output Menu
The OUTPUT menu provides functions for controlling the display and output
of data. The soft-key functions that appear in this menu are DISP WINDOW,
OUTPUT CHAN, PRINT OUTPUT, PRINT MEMORY,andCLEAR
STATS.