5.2.1.12 Thermocouple Polynomial Conversion
EDIT TCS 03
PROBE SER#: 1
CONVERSION: TC-POLY
CJC: INTERNAL
C0: 0.0
CJC TEMP: 0.0
C1: 0.038562
$
NOTE: An Application Note for use of Tungsten-Rhenium and other
thermocouples is available at www.hartscientific.com.
The following equation is used for the thermocouple polynomial conversion:
Et ct
i
i
i
( [ ])[ ]°=
=
∑
CVμ
1
15
NOTE: The EMF (E) in the above equation is in microvolts. If the cali-
bration equation for the coefficients of the probe being used is in milli-
volts, each coefficient needs to be multiplied by 1000 before entering the
coefficients into the instrument.
The user-defined parameters for the thermocouple polynomial conversion are
the coefficients C0 (c
0
) through C15 (c
15
). The user can specify internal or ex
-
ternal cold-junction compensation (CJC). If external CJC is selected, the user
can enter the temperature of the CJC reference. DC1 and DC2 are only accessi
-
ble when type R is used.
5.2.2 Copy Probe
The COPY PROBE soft-key can be used to copy an entire set of probe param
-
eters from one input channel to another. The parameters are copied in two
steps—select the source channel then select the destination channel. A window
appears requesting you to select the source channel. Each channel in the list is
identified with the device name, the channel number, and the probe serial num
-
ber. The list includes two USER channels that can be used for temporary stor
-
age. The source channel is selected using the UD buttons and pressing Enter.