8060A
Instruction Manual
5-12
5-9. Backup Fuse Replacement
Use the following procedure to replace the backup fuse (F2):
1. Remove the top cover by following the precautions given previously for
the calibration and backup fuse access.
2. Use a flat-tipped screwdriver to pry the fuse out of its fuse holder. Pry
the fuse from the side as indicated in Figure 5-1.
3. Replace the defective backup fuse with a 3A/600V type BBS-3 (Fluke
PN 475004). Refer to section 2-4 for information about replacing fuse
F1 (2A/250V; American style: fast acting type AGX2, 1/4 x 1”, Fluke
PN 376582.; European style: 5 x 20 mm, Fluke PN 460972).
5-10. Cleaning
Clean the front panel and case with a damp cloth and mild detergent. Do not
use abrasives, solvents, or alcohol.
Warning
To avoid electrical shock, remove test leads and any
input signals before cleaning operation.