Instruction Manual
5-4
Table 5-1. Required Test Equipment
Equipment Required specifications
Recommended
Type
DMM
Calibrator
DC Voltage: 0 to 1000V, ±(0.0075%)
AC Voltage:
200 Hz to 1 kHz, 0 to 750V, ±(0.06%)
1 kHz to 10 kHz, 0 to 200 V, ±(0.06%)
10 kHz to 30 kHz, 0 to 200V, ±(0.1%)
30 kHz to 50 kHz, 0 to 200V, ±(0.25%)
50 kHz to 100 kHz, 0 to 2.0V, ±(0.75%)
Resistance: 100 to 10.0 M, ±(0.1%)
Fluke 5100B
with Options
Y5000, 5100A-
03, and Fluke
5205A Amplifier
DC Current: 0 to 2000 mA, ±(0.05%)
AC Current:
20 Hz to 3 kHz, 0 to 2000 mA, ±(0.25%)
Fluke 5100B
with Option
Y5000 and
Fluke 5220A
Amplifier
Reference
Resistors
40 M and 290 M, ±(0.1%)
Caddock
MG750*
Signal
Source
Frequency: 25 mV to 200 mV, 100 Hz to
200 kHz, ±(0.1%)
Fluke 5700A
DMM
DC Voltage: 200 mV to 20V, ±(0.25%)
DC Current: 2 mA to 200 mA, ±(0.1%)
Fluke 87
*Precision high M resistors may be ordered from Caddock Electronics,
3127 Chicago Ave., Riverside, CA, 92507. Be sure to specify 0.1%
tolerance.
5-3. General Information
It is recommended that you periodically check the battery and perform the
performance tests (paragraphs 5-11 through 5-20).