8060A
Instruction Manual
5-18
Table 5-4. Current Test
Step Function Range
Input Signal
Display
Level Freq.
Limits
1
2
3
4
5
6
7
Current dc
200 µA
200 µA
2 mA
20 mA
200 mA
2000 mA
2000 mA
190.00 µA
-190.00 µA
1.9000 mA
19.000 mA
190.00 mA
1900.0 mA
-1900.0 mA
dc
189.60 to 190.40
-189.60 to -190.40
1.8960 to 1.9040
18.941 to 19.059
189.41 to 190.59
1894.1 to 1905.9
-1894.1 to -1905.9
8 Current ac 20 mA rms 19.000 mA 1 kHz 18.848 to 19.152
5-19. Diode Test
Use the following procedure to verify proper operation of the diode test:
1. Remove any connections to the inputs of the UUT and select the diode
test function. The display should indicate OL (overrange).
2. Connect the UUT and the DMM Calibrator as shown in Figure 5-4 and
apply a resistance of 1.000 k. a reading of .9000 to 1.1000 (typical)
should appear on the UUT display.
5-20. Frequency Test
To verify the proper operation of the frequency function, connect the signal
source to the V and COMMON inputs of the UUT. Select ac voltage and
push the Hz button to select the frequency function. Follow the steps listed in
Table 5-5.
Table 5-5. Frequency Test
AC Volts
Input Signal
Display Limits
Range
Level Freq.
200 mV
200 mV
2V
20.0 mV
100.0 mV
200.0 mV
100.00 Hz
190.00 kHz
19.000 kHz
99.94 to 100.06 ±3 counts*
189.91 to 190.09
18.991 to 19.009
*May be noisy due to low signal level.