Maintenance
Performance Tests
5
5-17
3. Apply a resistance of 1.000 M
. A reading of 993.0 to 1007.0 should
appear on the display.
4. Remove the resistance from the inputs to the UUT. The display should
read 0.0 to 0.2.
5-18. Current Test
Use the following procedures to verify proper operation of the dc and ac
current functions:
1. Connect the UUT and the DMM Calibrator as shown in Figure 5-5.
2. For each step in Table 5-4, select the UUT function and range as
indicated. Program the DMM Calibrator for the specified input signal
and verify that the displayed UUT value is within the indicated limits.
2000mA
2000nS
A
A
COMMON
V S
V
S
200mA
200
200k
200
µ
A
DC
AC
Hz
200mV
200
20mA
20
20k
2mA
2
2k
1000 DC
750 AC
M
Hz
dB
REL
1000V DC
750V AC
MAX
2A MAX
500V MAX
!
!
DMM
Calibrator
HI
UUT
LO
dx36f.eps
Figure 5-5. Equipment Connection for Current Test