EasyManua.ls Logo

Fluke 8060A - Autoranging Megohms

Fluke 8060A
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions
Operation
2
2-19
Resistance measurements for all ranges are made using a two-wire
ratiometric technique. This means that test lead resistance may affect the
accuracy in the 200
range. You can correct this error by shorting the test
leads together, reading the test lead resistance, and then subtracting it from
resistance readings. The most convenient way to do this is with the relative
function as described later in this chapter. This technique is also useful for
removing the 0.02
error factor in the 200
range (refer to resistance
specifications in Chapter 1).
2-18. Autoranging Megohms
When the autoranging M
range is selected, the 8060A automatically selects
the range appropriate for the measurement. The measurement resolution
decreases in the two higher M
ranges as shown in Table 2-3. Readings
made at the crossover points between ranges are microcomputer-stabilized
by an offset in the upscale and downscale directions. Range changes are
made at 2.00 M
and 20.00 M
as readings go upscale, or at 19.0 M
and
1.90 M as readings go downscale.
Table 2-3. Resistance Function Autoranges and Resolution
Range Resolution
No. of Digits
Possible in
Reading
2 M 100
M
20 M 10 k
Autorange
300 M
20 to 99.9 k
100 to 300 M
100 k
1 M
3
3
k
2 k
20 k
0.1
10
Autorange
300 k
20 to 99.9 k
100 to 299 k
100
1 k
3
3

Table of Contents

Other manuals for Fluke 8060A

Related product manuals