Instruction Manual
5-14
Table 5-2. Voltage Test
Step Function Range
Input Signal
Display
Level Freq.
Limits
1
2
3
4
5
6
DC Voltage 200 mV
200 mV
2V
20V
200V
1000V
+ 190.00 mV
-190.00 mV
1.9000V
19.000V
190.00V
1000.0V
dc 189.91 to 190.09
-189.91 to -190.09
1.8991 to 1.9009
18.989 to 19.011
189.89 to 190.11
999.3 to 1000.7
7
8
AC Voltage
AC Voltage
200 mV 100.00 mV rms
100.00 mV rms
50 Hz*
10 kHz
99.68 to 100.32
99.60 to 100.40
9 AC dB 2V 1.9000V rms 50 Hz* 7.69 to 7.89 (dB)
10
11
12
13
14
15
16
17
AC Voltage 2V 1.0000V rms
1.0000V rms
1.0000V rms
1.0000V rms
1.0000V rms
1.0000V rms
200.0 mV rms
200.0 mV rms
50 Hz*
1 kHz
10 kHz
30 kHz
100 kHz
20 Hz
50 Hz*
30 kHz
.9938 to 1.0062
.9938 to 1.0062
.9930 to 1.0070
.9860 to 1.0140
.9500 to 1.0500
.9890 to 1.0110
.1978 to .2022
.1940 to .2060
18
19
20
AC Voltage 20V 10.000V rms 50 Hz*
10 kHz
30 kHz
9.938 to 10.062
9.930 to 10.070
9.860 to 10.140
21
22
23
24
25
26
27
AC Voltage 20V
200V
200V
200V
200V
1000V
1000V
10.000V rms
100.00V rms
100.00V rms
100.00V rms
100.00V rms
750.0V rms
750.0V rms
100 kHz
50 Hz*
10 kHz
30 kHz
100 kHz
100 Hz
1 kHz
9.500 to 10.500
99.38 to 100.62
99.30 to 100.70
98.60 to 101.40
95.00 to 105.00
741.3 to 758.7
741.3 to 758.7
* If power line frequency is 50 Hz use 60 Hz test frequency.