8060A
Instruction Manual
1-14
Table 1-1. 8060A Specifications (cont)
General
Maximum Common
Mode Voltage............... 500V dc or ac rms
Display Update Rate ... 2.5 readings/second for all functions
except frequency and dB. For frequency,
1 reading/second. For dB, 1.4
readings/second.
Electromagnetic
Compatibility ..................
In an RF field of 1 V/m on all ranges
and functions: Total Accuracy =
Specified Accuracy + 2.3% of range.
Performance above 1 V/m is not
specified.
Display ............................ digit duplex LCD (19,999 counts),
leading zero suppression, autopolarity.
Display Annunciators .... BT (low battery indicator), Hz or kHz
(frequency unit), dB (dB function
enabled), REL (relative function
enabled).
and (continuity
function enabled), and (bar indicates
continuity detected).
A/D Converter................. Dual-slope converter
Power .............................. Single standard 9V battery (NEDA
1604)
Battery Life ..................... Typically 170 hours with an alkaline
battery. BT appears on display when
approximately 20% of battery life
remains.
Size.................................. 180 mm L x 86 mm W x 45 mm H (7.1”
L x 3.4” W x 1.8” H)
Weight ............................. .41 kg (14.5 oz.)
Shock and Vibration ...... MIL-T-28800B