Operating Instructions
Initial Check-Out Procedure
2
2-35
Push the
button twice to enable the visible and audible
continuity. You should see the bar in the display and hear the tone.
Select the diode test (with the V
S and A inputs still shorted together).
The display should read .0102
±
.0015V.
Select the conductance function (with the V
S and A inputs still
shorted together). The instrument should indicate overrange (OL).
Remove the connection between the inputs. The instrument should
indicate 0.0
±
1.0.