Vertical flanged machines Static clock
Use Static measurement mode
l
Center the laser beam.
Note
Static measurement mode is used for vertically mounted machines.
l
Rotate the shafts to the first measurement position. If the coupling housing numbering
convention is used, the reference mark on the housing and the measurement position 0
must be aligned.
l
Use or to position the displayed sensor and reflector at the same rota-
tional position as the actual components mounted on the shafts, then tap M [1] or
to take the first measurement point.
l
Rotate shaft to the second measurement position (e.g. 1:30). If the chosen meas-
urement position does not correspond to the angle selected automatically on the display,
use the navigation keys to manually position the sensor and reflector at correct position
on the display.
Tap M [1] to take measurement point.
l
Take the maximum number of measurement points to maximize the quality of results.