Automatic Input Terminal Detection
For ac and dc current measurement functions, the Meter automatically detects the signal
input between the mA and 10 A input terminals. A front-panel annunciator indicates
whether the Meter is in the mA or A ranges.
If a test lead is inserted in to the mA input terminal and there isn’t a test lead in the 10 A
terminal, then only the 200 µA to 200 mA ranges are selectable. If the 10 A terminal has
a test lead, then only the 2 A and 10 A ranges are selectable.