Operating the Meter from the Front Panel
Power-Up Configuration 3
3-25
Range mode on primary display (manual or autorange)
Measurement rate (slow, medium, fast)
Dual display status (active or inactive)
Any combination of selected function modifiers
Touch Hold level (1, 2, 3, 4)
Last recorded minimum and maximum values for MINMAX modifier
Last recorded relative base
Relative base shown in secondary display (enabled or disabled)
Last HI-LO settings in compare mode
Trigger mode (1, 2, 3, 4, 5)
Echo setting (on or off)
dB and dB reference
RS-232 settings
PRINT mode
Data format (with or without UNIT) sending through the RS-232
Power-Up Configuration
When the Meter is turned on and the power-up sequence is complete, the Meter defaults
to the power-up configuration listed in Table 3-9.
The RS-232 baud rate, parity, and echo mode are not changed when power is cycled off
and on. These parameters remain as set until changed by the user.
Table 3-9. Factory Power-Up Configuration
Parameter Configuration
Function setting DC volts
Range mode Autorange
Reading rate Slow (2.5 readings/second)
Touch Hold sensitivity level 1 (5 % of reading)
High/low values for Compare mode (COMP) 0
Minimum and maximum values in MIN MAX modifier 0
Relative base 0
Relative base in secondary display Disabled
Trigger type 1 (Internal)
Trigger type 0
Calibration
Refer to the 8808A Calibration Manual for instructions on calibrating the Meter.