EasyManuals Logo

Fluke 8808A User Manual

Fluke 8808A
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
8808A
Users Manual
3-20
Level 1 (5 % of range)
Level 2 (7 % of range)
Level 3 (8 % of range)
To change the response level, press Q and I. The response level currently selected
(1, 2, 3 or 4) appears on the primary display. Press U or V to step to the desired
response level, then press R for two seconds to set the level and return to the primary
display. You can return to the primary display without changing the response level by
pressing any button except R, U or V.
Minimum / Maximum Modifier (MIN MAX)
The minimum/maximum (MIN MAX) modifier stores the minimum and maximum
inputs of your measurements.
Selecting the MIN MAX modifier turns off autoranging and locks in the present range, so
make sure you are in the correct range before pressing N. If you press U or V
after pressing N, the Meter exits the MIN MAX modifier mode.
To store the minimum and maximum inputs:
1. Press N to select the MIN MAX modifier.
When N is first pressed, the minimum and maximum values are set to the
reading displayed. MAX is shown and the display indicates the latest maximum
reading.
2. Press N again to display the minimum reading. MIN is shown and the display
indicates the latest minimum reading.
3. Press N again to display either the minimum or maximum reading. M
is shown and the display indicates either the minimum or maximum reading.
4. To exit MIN MAX mode, press and hold N for 2 seconds.
5. To observe the actual readings without resetting the stored values, press Q and
then select the same measurement function that you selected for the primary
display.
Shop for Fluke products online at:
1.877.766.5412
www.MyFlukeStore.ca

Table of Contents

Other manuals for Fluke 8808A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 8808A and is the answer not in the manual?

Fluke 8808A Specifications

General IconGeneral
Product typeDigital multimeter
Product colorGrey, White
Frequency range0.2 - 2 kHz
AC current range0.2 - 10 A
AC voltage range0.2 - 750 V
DC current range0.0002 - 10 A
DC voltage range0.2 - 1000 V
Resistance range200 - 100000000 Ω
Placement supportedBench multimeter
Resistance resolution0.001 Ω
Resolution (frequency)0.0001 Hz
Resolution (AC current)100 µA
Resolution (DC current)1 µA
Resolution (DC voltage)1 mV
Basic accuracy (frequency)0.01%
Basic accuracy (DC current)0.02 + 0.005
Basic accuracy (DC voltage)0.015 + 0.003
Basic accuracy (resistance)0.02 + 0.003
Display digits5.5 digits
Weight and Dimensions IconWeight and Dimensions
Depth297 mm
Width217 mm
Height88 mm
Weight2100 g

Related product manuals