8808A 
Users Manual 
A-2 
 
 
eue25.eps 
Figure A-1. Example of Dual Display Showing Volts AC and Frequency 
Using Measurement Functions in Combination 
The dual display allows you to display select combinations of measurements for the input 
signal being measured. Allowable combinations of properties are shown in Table A-1. 
Volts (dc + ac) rms or current (dc + ac) rms measurements can only be made in the 
primary display. While (dc + ac) measurements are being made, another function cannot 
be selected for the secondary display. 
Additional combinations of dual readings can be added by using the relative readings, 
minimum/maximum, and/or Touch Hold function modifiers. 
Table A-1. Allowable Combinations of Measurements 
Primary Function  
DC V  AC V  DC I
[1]
 AC I  FREQ OHMS 
DC V  X X X X     
AC V  X X X X  X   
DC I  X X X X     
AC I  X X X X     
FREQ   X     X   
Secondary Function 
OHMS      X 
How the Meter Takes Dual Display Measurements 
When the Meter is in the dual display mode (both the primary and secondary displays are 
on), the Meter takes measurements and updates the displays in one of two ways: (1) It 
takes a single measurement and updates both displays using that measurement; or (2) it 
updates each display using a separate measurement. 
Updating Primary and Secondary Displays with a Single Measurement 
The Meter takes a measurement and updates both displays using that measurement only 
when the measurement function is the same for both the primary and secondary displays. 
This will happen, for instance, if Touch Hold (with autoranging on) is applied to a 
measurement function on the primary display and the same function is selected for the 
secondary display. 