CHANGING TO A NEW TYPE OF UUT
After you test or troubleshoot one type of UUT and are ready for
another type, you will typically follow the procedure shown
below. Section 4 describes self-tests, configuration, and
calibration.
1.
Perform self-tests on any new pod, probe, or
I/O
module added to the system.
2.
Change user disks if the new UUT directory is on a
different user disk.
3.
Recalibrate the system if the new UUT uses a new
pod or extra
I/O
modules.
4. Restore the configuration for the new UUT from the
user disk.
CHANGING
9100A/9105A
SYSTEM DEVICES 1.7.
If parts of the system such as the pod, probe, or
J/O
modules are
changed, you should follow the procedure below. Section 4
describes self-tests, configuration, and calibration.
1.
Perform self-tests for the new devices.
2.
Recalibrate the system.