Restoring Calibration Data
To restore calibration data from a user disk:
4.5.2.
-
1.
2.
3.
4.
5.
6.
Press the SETUP MENU key and use the left arrow
key to move the cursor to the left-most field.
Press the RESTORE
softkey.
Move the cursor to the next field and press the
CALDATA
softkey.
Move the cursor to the next field and press one of the
following softkeys:
USERDISK: If the calibration data was saved in the
USERDISK
directory. The resulting display should
read:
RESTORE
CALDATA
FROM
USERDISK
UUT FILE: If the calibration data was saved in a
UUT directory.
--
If you pressed the UUT FILE
softkey
in step 4, type
the UUT directory name; for example, type
“DEMO”. The resulting display should read:
RESTORE
CALDATA
FROM UUT FILE DEMO
Press ENTER to restore the previously saved
calibration data.
-
4-10