Fluke 9100A - No Pod Connected Error; Bad UUT Power Error

Fluke 9100A
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NO POD CONNECTED
3.1.
The pod is not connected to the system. To clear the error:
1.
Switch off UUT power.
-
2.
Make sure the pod is connected to the system.
3.
Connect the pod to the UUT.
4. Switch on UUT power and restart the test.
If the test fails again,
the pod may be defective (see
Section 4.2.2).
BAD UUT POWER
3.2.
The system detects that the UUT power supply is producing an
incorrect voltage. To clear the error:
1.
Switch off UUT power.
-
2. Make sure the pod is correctly connected to the UUT
test-access socket. For details, refer to the pod
manual.
3.
Check power connections to the UUT.
4. Switch on UUT power and restart the test.
If the test fails again, the UUT power supply is probably
defective.