After each power-up or reset before you test or trouble-
shoot a UUT:
0
Calibrate and configure the system for the type of UUT
you are about to test or troubleshoot.
l
Ensure that self-tests have been performed on the pod,
probe, and I/O modules.
0
Verify that the system mode is appropriate and that your
user disk is in the correct drive.
If you reset or change any part of the system:
l
The self-tests will have to be repeated and the system
recalibrated. You can save new calibration data on the user
disk, writing over the old data.
0
Test and troubleshooting programs and the system config-
uration remain valid.
If you change
UUTs
from one type to another:
l
You may have to change pods or I/O modules. If so,
repeat the self-tests and calibrations with the new pod.
l
System settings are probably different for the new UUT.
Restore settings from the new UUT directory to recon-
figure the system.