NOTE
It is recommended that your 9IOOA has at
least 3 megabytes of memory when using
the E-disk.
7. Press the ENTER/YES key. This creates the E-disk
and loads the UUT information into it.
An error message may appear on the
display;“E-Disk
is too
small to hold UUT -- make E-Disk larger”. Use the following
steps to increase the amount of RAM for the E-disk:
1.
Press the UP-ARROW key to clear the message off
the screen and enter a higher amount of RAM into the
right-most field.
2.
Press the ENTER/YES key to attempt loading again,
and repeat this procedure until you are able to load
the E-disk successfully.
During GFI execution and after E-disk has been loaded; if you
receive “Out of Memory” errors, there is not enough RAM to
run your programs and use the E-disk. You can delete E-disk
using the following steps:
1.
Press the MAIN MENU key.
2. Locate and press the
softkey
under E-DISK (you
may need to press the SOFT KEYS button to have
E-DISK appear on the display).
3.
Press the RIGHT ARROW key. The highlighted
word on the first line of the display should be
DELETE. If it is LOAD, press the
softkey
under
DELETE.
4. Press the ENTER/YES key.
The E-disk is now deleted, and the RAM used to store the
E-disk is returned for use by GFI.