Getting Started
Configuring the Test Tool
2
2-37
Storing Plot Data with Saved Autotest Results (DSP-4100)
When STORE PLOT DATA is enabled, saved Autotest results include plot data
from tests such as the attenuation and NEXT tests. HDTDX and HDTDR plots are
also saved with the Autotest results. Saving the plot data lets you include plots on
test reports uploaded to a PC and printed with CableManager Software. The DSP-
4100 can save the results of at least 250 Autotests when plot data is included.
More results can be saved if plot data is not included.
Note
Storing plot data with Autotest results reduces the number of results
you can save.
Enable or disable this setting as follows:
1. Turn the rotary switch to SETUP.
2. Use $ Page Down and D to find and highlight the store plot data
setting.
3. Press ! Choice.
4. Use D U to highlight the desired setting; then press E.
Selecting a Length Unit
The test tool displays length measurements in meters or feet.
To change the unit of measurement, proceed as follows:
1. Turn the rotary switch to SETUP.
2. Press $ Page Down until you see the length units setting.
3. Press ! Choice.
4. Use D U to highlight the desired unit.
5. Press E to accept the highlighted unit.