Step Action
8 If your system is equipped with a second pump and the gradient setup, re-
p
eat the the procedure from step one for the inlets on the B line.
Tip:
Inlet B4 is the inlet located furthest from the pump, Inlet B3 is the second
furthest etc.
Inlet B1 is the recommended inlet to use for prime flow as it leads straight
to the pump.
9 Select E
ND to stop the pump and close the valves.
Result:
The air has been removed from the flow path.
5 Preparation
5
.4 Prime to remove air from the pump and flow path
128
ÄKTA ready XL Operating Instructions 29281616 AE
*** TeDIS Valid on Date Printed Printed on: 15-09-2023 11:27:35 ***