Step Action
2 Select the method to start. Click O
K.
A Start Protocol is displayed, consisting of a number of dialog boxes. For
example, on the Variables page, it is possible to fine-tune the method be-
fore proceeding. Selecting the Show details checkbox will display more
detailed information.
3 Click N
ext or Back to navigate through the dialog pages, adding the infor-
mation required as well as your own comments.
4 Click START in the Result Name dialog box. This will initiate the method
run.
Monitor a run
T
he progress of the run can be viewed in detail in the UNICORN System Control
module.
Up to four different panels can simultaneously display different aspects of the run. To
customize content and layout of a panel, right-click and then click Customize. By
clicking Documentation and then Run notes, it is possible to add notes during a run.
Entered notes will be included in the results.
It is possible to set flow control measured from the flow meter or calculated from
pump RPM under Monitors. The function can be used for non-conducting liquids
such as ultra pure water of WFI. Refer to the UNICORN helptext for more details.
6 Operation
6
.2 Run a method
ÄKTA ready XL Operating Instructions 29281616 AE 135