Troubleshooting
2001989-203C ApexPro™ Telemetry 6-19
6. Select Stop Blink.
Blink rack
Start blinking a diagnostic LED for a receiver subsystem
To start blinking the system status LED of a particular receiver subsystem attached to
the specified tower, follow these steps.
1. Log onto the Webmin service interface on page 4-7.
2. Select the Diagnostics tab.
3. Select Receiver Subsystem > Blink Rack.
140A
4. In the Sort box, select the desired option for sorting the tower list.
5. Select the tower from the Tower drop-down list.
6. Select the number of minutes to blink the LED from the Minute drop-down list.
7. Select the IP address of the receiver subsystem you want to blink from the Rack
drop-down list.
8. Select Blink.
Stop blinking a diagnostic LED for a receiver subsystem
To stop blinking the system status LED of a particular receiver subsystem attached to
the specified tower, follow these steps.
1. Log onto the Webmin service interface on page 4-7.
2. Select the Diagnostics tab.
3. Select Receiver Subsystem > Blink Rack.
4. In the Sort box, select the desired option for sorting the tower list.
5. Select the tower from the Tower drop-down list.
6. Select Stop Blink.