Parts lists, drawings and replacement
2001989-203C ApexPro™ Telemetry 7-33
070A
NOTE
Note the position of the temperature sensor DIP switches for re-assembly.
7. Using a phillips screwdriver, remove the screw securing the temperature sensor
at the back of the chassis floor, next to the power supply module.
075A
NOTE
Note the position of the temperature sensor DIP switches for re-assembly.
8. Disconnect the temperature sensor wiring harness plug from the receptacle at the
back right edge of the SNMP.
076A
Number Position
1OFF
2OFF
3ON
4ON
Number Position
1OFF
2OFF
3OFF
4ON