Configuration
2001989-203C ApexPro™ Telemetry 5-11
Configure the Remote Service settings
To view and configure the Remote Service settings, perform the following steps.
1. Log onto the Webmin service interface on page 4-7.
2. Select the Configuration tab.
3. Select Remote Service > Configuration.
124A
4. On the Remote Service Configuration screen, make the necessary changes to the
appropriate fields:
NOTE
Remote Service Configuration settings are automatically populated.
HTTP Proxy Server Configuration
Parameter Value Notes
Address If this site uses an HTTP
proxy server, a specific IP
Address and Port number
are required for the Remote
Service communication to
work. Otherwise, select
None.
If the HTTP proxy server
requires user authorization,
a specific Username and
Password is required.
Otherwise, select None.
These values are determined
during the site survey. See the
site survey workbook.
When configuring the HTTP
Proxy Server, if you select
None, do not to enter any
information in the Address,
Port, Username, or
Password fields. If you select
None, the Webmin page will
not save the address, port,
username and password.
Port
Username and
Password

Table of Contents

Other manuals for GE ApexPro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ApexPro and is the answer not in the manual?

GE ApexPro Specifications

General IconGeneral
BrandGE
ModelApexPro
CategoryMedical Equipment
LanguageEnglish

Related product manuals