Chapter 7. Reference
90 | ADTS 405 R/F Mk2 Instruction Manual-English © 2017 General Electric Company – All Rights Reserved.
7.3.25 SETUP, or (Nudge Keys)
This function allows the increment value of the nudge key to be adjusted for every parameter.
1. When SETUP, or are pressed, the display shows the value of nudge for the
parameter last displayed.
2. Enter a new value to replace the present value.
3. Use [NEXT] or [PREV] to display other parameters.
4. Press CLEAR/QUIT when the nudge values of all the required parameters have been
changed.
In aeronautical units, the parameters available are:
•ALT
CAS/TAS
•MACH
ROC
RATE CAS
In pressure units, the parameters available are:
Ps, RATE Ps
•Pt
Qc, RATE Qc
EPR, RATE EPR
5. The nudge value can be used for large increments e.g. 100 kts or 5,000 ft to “step
through” the cardinal points of a test.
7.3.26 MINIMUM SETUP
Minimum setup allows access to a limited number of setup functions. Any parameter
changed under setup returns to the default settings at power-down. The CONFIG mode may
be used to change the default settings so that the ADTS 405 Mk2 powers-up in the required
state.
7.3.27 SETUP, [UNITS], [AERO]
The current operating limits are changed to the default aeronautical units. These units can be
programmed using CONFIG, [UNITS], [AERO].
7.3.28 SETUP, [UNITS], [PRESS]
The current operating limits are changed to the default pressure units. These units can be
programmed using CONFIG, [UNITS], [PRESS].
INFORMATION Make sure the nudge increment is set to a suitable value
before using the nudge facility to change the aim value.