CONFIGURATION
© 2017 General Electric Company – All Rights Reserved. ADTS 405 R/F Mk2 Instruction Manual-English | 93
3. The default [TEMPERATURE] units are used for entry of temperature values in SETUP
or CONFIG.
4. The default [AERONAUTICAL] and [PRESSURE] are set in MINIMUM SETUP, [UNITS].
7.4.4 CONFIG, [LIMITS], [EDIT LIMITS], [EDIT EXISTING]
This function allows existing sets of limits to be configured.
a. Use [NEXT] or [PREV] to select the set of limits by aircraft name or component/
system name/number when the display shows the required aircraft limits, press [SEL]
to edit the limit values.
b. Use [NEXT] or [PREV] to step through each limit or associated parameter in turn.
Enter a new numeric value to change a limit.
c. Press CLEAR/QUIT when all required changes have been made and select [YES] to
save changes.
The ADTS 405 Mk2, when delivered, contains “Standard” and “Civil” limits; to set and store
“Max” limits, for the first time, enter configuration. The “Max” limits depend on the range of
the pitot channel either 850 or 1000 kts, see Section 1.3, “Operating Limits,” on page 2. The
system's defined “STANDARD”, “CIVIL” and “MAX” limit sets cannot be redefined. A defined
set can be edited but must be saved using a different name. This method could save time.
The parameters are:
Name
Max Alt
Min Alt
Max CAS
Min CAS
Max Mach
Max ROC
Max Rate CAS
Max Ps
Min Ps
Max Qc
Min Qc
Max Rate Ps
Max Rate Qc
ARINC Limits
Altitude Correction
7.4.5 NAME
The aircraft name or component/system name/number is used for selecting the set of limits
in SETUP or CONFIG. Enter up to 8 characters using 0 to 9, - (minus), (decimal point) and any
alphabetic characters in any order or combination.