Chapter 1. Description
2 | ADTS 405 R/F Mk2 Instruction Manual-English © 2017 General Electric Company – All Rights Reserved.
Operating limits are set, predefined tabular limits known as STANDARD, CIVIL and MAX these
can be selected through the SETUP menu (see Section 7.3.3, “SETUP, [LIMITS],” on page 83).
Operators may also configure the display to aeronautical or pressure units but should be
aware that when units of pressure are selected, wider full-scale pressure limits will be
enabled for some parameters.
1.3 Operating Limits
The following sets of operating limits are supplied with the ADTS 405 Mk2.
Table 1-1: Operating Limits - 850 kts Range
Parameter Max Limits Standard Limits Civil Limits
Min Alt -3,000 ft -2,000 ft -1,000 ft
Max Alt 105,000 ft 80,000 ft 50,000 ft
Min CAS -100 kts 0.0 kts 0 kts
Max CAS 850 kts 850 kts 450 kts
Min Ps 3.0 mbar 27.615 mbar 115.972 mbar
Max Ps 1355.00 mbar 1088.657 mbar 1050.406 mbar
Min Qc -1,352.00 mbar 0.0 mbar 0.0 mbar
Max Qc 1,700.00 mbar 1688.00 mbar 368.01 mbar
Max Mach 12.35 2.5 1
Max ROC 100,000 ft/min 9,000 ft/min 6,000 ft/min
Max Rate CAS 2,000 kts/min 600 kts/min 600 kts/min
Max Rate Ps 10,000.00 mbar/min 200.00 mbar/min 100.00 mbar/min
Max Rate Qc 10,000.00 mbar/min 200.00 mbar/min 100.00 mbar/min
ARINC Limits OFF OFF OFF