Chapter 8. Specification
102 | ADTS 405 R/F Mk2 Instruction Manual-English © 2017 General Electric Company – All Rights Reserved.
Table 8-2: Flight-line ADTS 405F Mk2 Specification
Parameter Value
Size
a
a. Nominal value. Actual value may vary due to configured optional extras.
Height: 463 mm (18.22")
Length: 781 mm (30.73")
Width: 331 mm (13.03")
Weight
a
33.5 kg (73.9 lb)
EMC EN 61326-1
MIL-STD-461F (Extended
environment)
Electrical Safety EN 61010-1
AC Power Supply Single phase AC in the range
(automatic range switching):
100/120/230 V, 50/60 Hz
115 V, 400 Hz
Power: 500 VA
Installation Category: II
Fuse: T5AH250V
DC Power Supply Supply range: 16 to 30 V
Power: 500 VA
Fuse: T20AH250V (left)
T32AH250V (right)
Operating Environment Indoor and outdoor use.
Operating Altitude Up to 4,570 metres (15,000 ft)
Operating Temperature Standard Range -20°C to 50°C (-4°F to 122°F)
Extended Range -40°C to 55°C (-40°F to 131°F)
Storage and Shipping
Temperature
-51°C to 70°C (-60°F to 158°F)
Operating Humidity 0 to 95% RH (non-condensing)
Pollution Degree 3
Pressure Safety Pressure Equipment Directive Class: Sound Engineering Practice (SEP)
Shock and Vibration ETSI EN 300 019-2-5 Class 5.1
Ground Vehicle Installations. Test
method: EN 60068-2-64 Random
Vibration.
Pressure Fittings Ps: MS33656-6
Pt: MS33656-4