Girae Warmer   3-2
WARNING:
Radiant energy may cause more rapid urine evaporation, and may lead to inaccurate urine 
diagnostic test analysis and inaccurate weight measurements. Frequent diaper changes 
are recommended.
WARNING:
Use Baby Mode unless Manual Mode is specically prescribed. While both modes require 
patient monitoring, Manual Mode requires constant attention. In Manual Mode, you must 
take the responsibility for detecting changes in the environment (high air ow, direct sun-
light, phototherapy lamp usage, etc.) or the patient’s condition and make heater power 
adjustments in response to these changes. In Baby Mode, the warmer automatically adjusts 
heater output to maintain the desired skin temperature, reducing (but not eliminating) the 
need to monitor the patient and make adjustments to the Baby Mode control temperature.
WARNING:
Do not leave the baby unattended while any bedside panels are lowered or removed.
WARNING:
When bedside panels are in an upright position, ensure that the bedside panels are properly 
locked in place.
WARNING:
The patient probe is not isolated from earth ground. Any additional equipment used with 
the Girae Warmer must comply with IEC 601. 
WARNING:
Bed-to-heater spacing less than approximately 85 cm will result in incorrect operation and 
may adversely aect the patient’s condition.
WARNING:
Do not place objects in the radiant heat path.  Objects will be heated and could block heat 
to the baby.
WARNING:
Do not hang items from the heater head.
WARNING:
Do not connect unapproved equipment into accessory outlets.