3-3Girae Warmer
Operating the Warmer (continued)
Start up
Illustration
Step
Description
1
Plug the unit into a properly rated AC power
outlet and set the caster brakes.
2
Place the mains power switch, located by the
outlets on the back of the unit, in the On (I)
position.
3
Place the Stand By power switch, located
on the left side of the controller, in the On (I)
position. Check the control panel to ensure
the screen is working properly.
Note: If the unit has been used in the last 2
hours, the patient history query appears. If
you select “No” the warmer uses the previous
settings, and you may skip to “Attaching the
Skin Probe” later in this section. If you select
“Yes” go to step 4.