EasyManua.ls Logo

GE Healthcare Brivo XR118 - Page 123

GE Healthcare Brivo XR118
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 5 - - Installation & Configuration Page 123
11.1.2 Wireless Signal Strength Test
Go to SUIF -> Diagnostics -> Wireless Detector -> Signal Strength. Click the “Start” to test the
wireless communication quality.
If there is no bucky on the wall stand, just test the wireless signal with detector in the table
bucky. Or else, just test the wireless signal with detector in the wall stand. Read the wireless
signal after the AP is installed 30 seconds later to get the latest value.
Two wireless communication criteria are required for our system:
- Bandwidth >=15, and signal strength >= -75dbm
After the AP is placed at a location, wireless signal quality need be checked.
If “Signal strength” <= Middle, relocate the detector:
- a) Be sure no object is blocking detector’s antenna.
- b) Be sure the distance of detector and hardware AP is <= 4m.
If “Signal strength” is Good, but “Bandwidth” <= Bad.
- a) Change the working channel of hardware AP.