DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 128 Section 12.0 - Leakage Current Test
12.1.1 Personnel Requirements
12.1.2 Overview
This procedure can be performed with:
• Fluke tester
12.1.3 Tools and Test Equipment
Note: The equivalent means that the tool is designed to be capable of performing tests to both the
standard of IEC 62353 and the earth leakage current of IEC 60601-1.
RISK OF ELECTRIC SHOCK.
VOLTAGE PRESENT.
THE SYSTEM MUST BE DISCONNECTED FROM THE FACILITY MAINS INCLUDING THE A1
PANEL GROUND WIRE.
12.1.4 Setup the tools
Refer to Chapter 1 4.5.5 Set up tools on page 29
12.1.5 Leakage Current Test Procedure
• Earth Leakage Current Test
1) Connected the BRIVO XR118 power plug to tool output receptacle.
Personnel Requirements Preliminary Reqs Procedure Finalization
1 Not Applicable 30min 10min
Item Qty. Effectivity Part# Manufacturer
Fluke ESA612 or equivalent 1 - - Fluke
No TEST POINT Expected Value
(mA)
Measure Value
(mA)
RESULT
1 Connected the BRIVO XR118
power plug to tool output
receptacle
10